EasyManua.ls Logo

Applied Biosystems SimpliAmp Thermal Cycler - Page 68

Applied Biosystems SimpliAmp Thermal Cycler
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Category Run method
High Fidelity
• AccuPrime
™
PFX 2-Step Method
• AccuPrime
™
PFX 3-Step Method
• AccuPrime
™
Taq High Fidelity Method
• Platinum
®
PFX 2-Step Method
• Platinum
®
PFX 3-Step Method
• Platinum
®
Taq High Fidelity Method
High Specificity
• AccuPrime
™
PFX 2-Step Method
• AccuPrime
™
PFX 3-Step Method
• AccuPrime
™
Taq DNA Polymerase Method
• AccuPrime
™
Taq High Fidelity Method
Next-Gen
Sequencing
TargetSeq
™
Exome Enrichment Method
Long PCR
• Elongase
®
2-Step Method
• Elongase
®
3-Step Method
• AccuPrime
™
Taq High Fidelity Method
• Platinum
®
Taq High Fidelity Method
Digital PCR (only for
Dual Flat block)
• 3D
• OpenArray
™
Genotyping
These run methods cannot be edited or deleted, but if you save them with a dierent
name or to a dierent folder, you can edit them.
The following sections provide the method details:
Note: For illustration purpose, the tables display maximum ramp rate. These will
however change with dierent block types.
AccuPrime
™
PFX 2-Step Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 35 1 — 50 105.0
Step 1 RampRate 6.0 6.0 6.0 —
Temperature 94.0 94.0 4.0
Hold Time 120 15 Infinite
Step 2 RampRate — 6.0 — —
Temperature 68.0
Hold Time 60
Appendix E Predefined run methods
Location of power point and ports on the instrument
E
68
SimpliAmp
™
Thermal Cycler User Guide