AccuPrime
PFX 3-Step Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 25 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 94.0 94.0 4.0
Hold Time 120 15 Infinite
Step 2 RampRate 6.0
Temperature 55.0
Hold Time 15
Step 3 RampRate 6.0
Temperature 68.0
Hold Time 60
AccuPrime
Taq DNA Polymerase Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 25 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 94.0 94.0 4.0
Hold Time 120 15 Infinite
Step 2 RampRate 6.0
Temperature 55.0
Hold Time 15
Step 3 RampRate 6.0
Temperature 68.0
Hold Time 60