EasyManua.ls Logo

Applied Biosystems SimpliAmp Thermal Cycler - Page 72

Applied Biosystems SimpliAmp Thermal Cycler
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AmpliTaq Gold
Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 35 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 95.0 95.0 72.0
Hold Time 300 15 420
Step 2 RampRate 6.0 6.0
Temperature 55.0 4.0
Hold Time 15 Infinite
Step 3 RampRate 6.0
Temperature 72.0
Hold Time 30
AmpliTaq
Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 35 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 95.0 95.0 72.0
Hold Time 60 15 420
Step 2 RampRate 6.0 6.0
Temperature 55.0 4.0
Hold Time 15 Infinite
Step 3 RampRate 6.0
Temperature 72.0
Hold Time 30
Appendix E Predefined run methods
Location of power point and ports on the instrument
E
72
SimpliAmp
Thermal Cycler User Guide

Table of Contents