AmpliTaq Gold
Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 35 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 95.0 95.0 72.0
Hold Time 300 15 420
Step 2 RampRate 6.0 6.0
Temperature 55.0 4.0
Hold Time 15 Infinite
Step 3 RampRate 6.0
Temperature 72.0
Hold Time 30
AmpliTaq
Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 35 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 95.0 95.0 72.0
Hold Time 60 15 420
Step 2 RampRate 6.0 6.0
Temperature 55.0 4.0
Hold Time 15 Infinite
Step 3 RampRate 6.0
Temperature 72.0
Hold Time 30