BigDye
Kit Standard Run Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 25 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 96.0 96.0 4.0
Hold Time 60 10 Infinite
Step 2 RampRate 6.0
Temperature 50.0
Hold Time 5
Step 3 RampRate 6.0
Temperature 60.0
Hold Time 240
Elongase
®
2-Step Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 30 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 94.0 94.0 4.0
Hold Time 30 30 Infinite
Step 2 RampRate 6.0
Temperature 68.0
Hold Time 60
Elongase
®
3-Step Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 30 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 94.0 94.0 4.0
Hold Time 60 30 Infinite
Appendix E Predefined run methods