EasyManua.ls Logo

Applied Biosystems SimpliAmp Thermal Cycler - Page 77

Applied Biosystems SimpliAmp Thermal Cycler
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Step 2 Hold Time 60 50 105.0
Platinium
®
PFX 3-Step Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 35 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 94.0 94.0 4.0
Hold Time 120 15 Infinite
Step 2 RampRate 6.0
Temperature 55.0
Hold Time 30
Step 3 RampRate 6.0
Temperature 68.0
Hold Time 60
Platinium
®
Taq DNA Polymerase Method
Cycle/Step
Stage Reaction
volume
(µL)
Cover
temp. (°C)
1 2 3 4
Cycles 1 30 1 50 105.0
Step 1 RampRate 6.0 6.0 6.0
Temperature 94.0 94.0 4.0
Hold Time 30 30 Infinite
Step 2 RampRate 6.0
Temperature 55.0
Hold Time 30
Step 3 RampRate 6.0
Temperature 72.0
Hold Time 60
Appendix E Predefined run methods
Location of power point and ports on the instrument
E
SimpliAmp
Thermal Cycler User Guide
77