Safety
182099015-V01-10.Juni2013-GA-MCL31-THAI 5
2.2 Electrical hazards
WARNING
The laser is operated with mains power. There is a risk of electric shock! Do not open
the device unless you are trained and authorized to do so.
The device is only to be operated with reliably grounded and correctly installed sockets
with grounding contact.
WARNING
The laser is operated with mains power. There is the risk of electric shock!
The device is designed to prevent liquids penetrating it in normal use. If liquid
penetrates the inner system space or the handpiece in exceptional cases, you must
turn the power off and pull out the mains power plug. Please contact our Engineering
Department.
WARNING
The handpiece spacer is in contact the patient’s skin during the treatment. There is the
risk of micro-organisms being transferred.
The selected spacer must be deep cleaned and disinfected directly after each
application!
2.4 Radio interference
The device complies with the requirements of the EN 60601-1-2 standard. The system is not
affected by electromagnetic interference generated by other devices that conform to the same
standard. Moreover, the system does not generate electromagnetic radiation beyond the limiting
values indicated in EN 60601-1-2. The manufacturer’s electromagnetic compatibility declaration
according to EN 60601-1-2 is enclosed with this manual.
WARNING
Medical devices are subject to specific electromagnetic compatibility (EMC)
precautionary measures. The device emits low levels of electromagnetic radiation that
may have an effect on devices that do not comply with the standard. Unintended
settings at these devices could be the result of such an effect.
Please observe the notes in the manufacturer’s EMC declaration in the Appendix when
installing and operating the device.
WARNING
The device may be influenced by devices that do not comply with the standard
indicated above. The result may be unintended settings. Therefore, switch off
cellphones and similar equipment before operating the laser device.
WARNING
The use of accessories and cables not authorized by the manufacturer may result in
increased interference emissions or decreased immunity of the device with the above
mentioned consequences.
Use only accessories and cables that have been authorized by Asclepion Laser
Technologies for the purpose