EasyManua.ls Logo

Asclepion MCL 31 Dermablate - Technical data; Device models; Specifications

Asclepion MCL 31 Dermablate
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data
182099015-V01-10.Juni2013-GA-MCL31-THAI 9
3 Technical data
3.1 Device models
Designation
Part
Number
Description
MCL 31 Dermablate
1820
Er:YAG Laser (main unit)
VarioTEAM Handpiece
4075
Standard handpiece for skin ablation (1-6mm)
VarioTEAM XL Handpiece
4076
Standard handpiece for skin ablation (8-12 mm)
MicroSpot Zoom Handpiece
4077
Handpiece for fractional skin resurfacing
The standard delivery includes the main unit with VarioTEAM Handpiece.
The VarioTEAM XL Handpiece and the MicroSpot Zoom Handpiece can be obtained optionally.
3.2 Specifications
Specification
MCL 31 Dermablate
Device model
Floor-based unit
Display
Display with capacitive push-buttons below glass plate
Operator guidance
Touchscreen
Cooling
Coolant
Internal cooling cycle
Distilled water only
Door interlock connector
5V/10mA
Laser warning lamp
Potential-free relay contact
max. 24V/1A (make contact)
Permissible ambient
conditions
Temperature: 15°C to 30°C
Rel. air humidity: max. 85% (no condensation)
Height: max. 2000m above sea level
Permissible transport and
storage conditions
Temperature: 10°C to +70°C
Rel. air humidity: max. 90% (no condensation)
Dimensions
360 × 600 × 930 mm (W × D × H)
without mirror arm and handpiece
Weight
approx. 75 kg
Power requirements
230V/50Hz max. 2500W
120V/60Hz max. 2500W
Main fuse
overcurrent release 16A, medium slow-blow
Overvoltage category
II (IEC 60664-1)
Peak mains voltage
Max. nominal mains voltage
V MT peak = 2500V peak
V MN rms = 300V rms