EasyManua.ls Logo

Asclepion MCL 31 Dermablate - Switching on

Asclepion MCL 31 Dermablate
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Start up
28 182099015-V01-10.Juni2013-GA-MCL31-THAI
4.3 Switching on
Verify that the following requirements have been met before switching on the device:
The device is in perfect condition (no damage).
The voltage specified on the type label is the same as that of the mains supply available
at the installation location.
Its power plug has been plugged into an appropriate power outlet.
The foot switch plug has been properly connected to the foot switch socket at the rear
panel of the device and screwed down.
The plug of the door contact is properly connected and screwed to its terminal at the rear
panel of the device.
The handpiece is connected.
Warning lamps are switched on at every laser room entrance.
Make sure that all safety precautions have been taken.
Then, proceed as follows:
1. Switch the power switch at the rear panel of the device to I (On).
(Figure 6: Power switch at the rear panel)
2. Verify that the Laser emergency STOP is deactivated (not pressed)
3. Now put the safety key into the key switch and turn it clockwise as far as it will
go.
While the device is automatically testing essential and safety-related parts, the start screen
appears on the display. The main menu appears on the screen.