Cleaning, disinfection and sterilization
182099015-V01-10.Juni2013-GA-MCL31-THAI 51
8 Cleaning, disinfection and sterilization
8.1 General Rules
Device
WARNING
The laser is operated with the mains power supply. There is the risk of electric shock!
Disconnect the mains lead from the mains socket to break the power supply to the
system before you proceed with cleaning.
Only outer system surfaces may be cleaned by the user.
WARNING
Only outer system surfaces may be cleaned by the user..
All accessible system components may be cleaned using a soft, slightly moistened piece of cloth.
Under no circumstance may a wet or dripping-wet piece of cloth be used for this purpose.
To remove stubborn contamination, a soft cleaner solution or disinfectant may be applied. Refrain
from the use of aggressive disinfectants or scouring agents.
Handpieces
WARNING
The contact surface of the distance holder of the handpiece contacts the patient’s skin
during the treatment. There is the risk of microorganisms being transferred.
Handpiece parts must be cleaned and disinfected immediately following use and before
application on a patient.
If it should be necessary to clean the device during a treatment the device must be
switched to the STANDBY mode before starting cleaning! Do not use alcohol or
isopropyl alcohol near the point of treatment because this causes a risk of fire if the
laser is activated afterwards!
CAUTION
Dirt settling on an optical surface may burn in when exposed to laser radiation and
cause side effects such as reddening, possibly blistering, incrustation, seldom scars!
Ensure that all accessible optical surfaces are undamaged and clean after the cleaning
process and remain clean during the treatment.
The procedure instructions given in the following sections are to be followed.
Important information
You are prohibited from cleaning a handpiece basic body under running water. This
may destroy the internal optics and electronics.
The handpieces cannot be sterilized.
If the handpieces are not in use for a long period they must be protected against
contamination.
CAUTION
It is possible to sterilize the spacer of the laser handpiece. The decision, if sterilization
is necessary, depends on the kind of treatment and is the responsibility of the doctor.
Handpieces must be protected from contamination if they are not required for some
time.