Start up
24 182099015-V01-10.Juni2013-GA-MCL31-THAI
4.2.3 Overview of the installation
Important information
After installation and any relocation of the device from a cold to a warm environment
with a temperature difference of more than 5°C, allow the device to adjust to the room
temperature in the unpacked state before being used for at least the following times:
min. 2 hours at a temperature difference of up to 10°C
min. 4 hours at a temperature difference of up to 15°C
min. 8 hours at a temperature difference of more than 20°C.
Afterwards, switch the device on without a handpiece connected and allow the device
to warm up for at least 30 minutes.
Take care that the installation and operating environment of the device meets the
following conditions: temperature: 15 to 30°C, relative humidity: max. 85% and height
not more than 2000 m above sea level.
Failure to observe these instructions may result in the destruction of the device.
For efficient cooling of the device, maintain a minimum separation of at least 20 cm between its
lateral and rear sides from the wall.
WARNING
The laser device emits high levels of radiation in the infrared range that may cause
serious damage to the eyes if the device is used incorrectly.
Every entrance to the laser area (usually the laser treatment room) must be equipped
with warning lamps that are lighting (or flashing) for as long as the laser is emitting
radiation to warn persons against entering the room.
WARNING
Make sure that the device cannot be moved unintentionally by locking its wheels. You
should also check that no part of the instrument has been damaged. To avoid injuries,
damaged parts must be replaced immediately!! While a part of the instrument remains
damaged, you must not use it – this is mandatory!!
WARNING
Make sure that there is enough space between walls and laser air openings to allow for
efficient air cooling. Do not cover these openings.