Setting the Alarm Volume7-8.
The alarm volume can be adjusted as desired.
Press for one second while pressing when the
power is on, and the current setting of the alarm volume
will appear on the setting display.
The alarm volume is adjustable in 8 levels from “ ”
(Minimum) to “ ” (Maximum).
Press or to set the alarm volume to a desired
level.
Changing the Temperature Deviation Limit of the Set 7-9.
Temperature Alarm
The set temperature alarm will occur if the infant’s skin
temperature (detected by the skin temperature 1) deviates
from the set temperature in the servo control mode. The
temperature deviation limit at which the set temperature
alarm will occur can be changed as follows.
Press for one second while pressing , and the
current setting of the temperature deviation limit will
appear on the setting display.
“ ”: set temperature ±1.0°C (±1.8°F)
“ ”: set temperature ±0.5°C (±0.9°F)
Press or to select a desired setting.