Inspection Checklist9-4.
WARNING
If any trouble is detected in any inspection, indicate on the unit that it is out of order and seek repair
without fail.
Contact your local Atom representative for further information on repairs.
Serial No. ( ) Date of Inspection ( . . ) Inspector ( )
No. Item to check Judgment
Inspection before use (1) (before connecting the power cord)
1 Is the main body free of breakage and other defects? YES/NO
2 Is the power cord free of damage and other defects? YES/NO
3 Are the power connections clean with no medical fluid stains? YES/NO
4 Do the casters rotate smoothly? Are they locked reliably? YES/NO
5
Are the baby guards attached to the mattress platform securely? Are the hinges of the baby
guards free of damage?
YES/NO
6 Is the mattress platform locked reliably in a tilted position? YES/NO
7 Is the area around the skin temperature probe connecting ports intact and clean? YES/NO
8 Are the control knobs attached securely? Do they operate reliably? YES/NO
9 Is the reflector intact and clean? YES/NO
10 Is the canopy held in place securely when tilted? YES/NO
11 Is the canopy fixing screw attached securely where it should be? YES/NO
12
Does the power failure alarm activate when the power switch is turned on? (The power failure
alarm indicator should come on and an audible alarm should sound.)
YES/NO
13 Are the tube introduction slit packings free of breakage and attached reliably? YES/NO
14 Are all the caution/warning stickers affixed securely? YES/NO
15 Is the Operation Manual kept where it is readily accessible? YES/NO
Inspection before use (2) (after connecting the power cord)
1 Does the mains on indicator come on when the power cord is connected to the power outlet? YES/NO
2 Does the height adjustment mechanism operate smoothly without making a strange sound? YES/NO
3
Do all the indicators on the operation panel, except for the power failure alarm indicator and
the battery operation indicator, come on and does an audible alarm sound when the power
switch is turned on?
YES/NO
4 Is an appropriate temperature displayed when the skin temperature probe is connected? YES/NO
5 Are SpO
2 and pulse rate values displayed when the SpO2 sensor is attached? YES/NO
6 Does each switch function properly when operated according to the Operation Manual? YES/NO
7
Does intensity of the lighting lamp increase gradually as the light control knob is turned
clockwise? Does the lighting lamp go out when the light control knob is turned fully counter-
clockwise?
YES/NO
Quarterly inspection (each time)
1
Does the heater output decrease when the skin temperature probe is warmed? Does the number
of bars on the heater output indicator change when the skin temperature probe is cooled?
YES/NO
2
Does the set temperature alarm indicator come on and does an audible alarm sound when the
skin temperature detected by the skin temperature probe deviates from the set temperature
±1.0°C or the set temperature ±0.5°C?
YES/NO
3
Does the set temperature alarm indicator flash and does an audible alarm sound when the skin
temperature detected by the skin temperature probe is 40°C or higher?
YES/NO
4 Does the baby check alarm indicator come on and does an audible alarm sound in 15 minutes? YES/NO