Operation Panel2-3.
Timer
Pulse oximeter
Alarm
Others
(Masimo) (Nellcor)
Temperature
External alarm indicator
Comes on when a high priority alarm
occurs, except for a power failure
alarm.
Mains on indicator
Comes on when the power cord is
connected to the AC power source.
Battery operation indicator
Comes on when the optional Power
Pack i (UPS: option) is being used as
a power source.
Power failure alarm indicator
Comes on when the power cord gets
disconnected from the AC power
source while the unit is in operation
or in the event of power failure.
System failure alarm indicator
Flashes when a system error or some
other trouble is detected.
Alarm silence indicator
Comes on when an audible alarm is
silenced.
Alarm
Function
Symbol
Set temperature alarm indicator
Comes on when a displayed skin
temperature deviates from the set
temperature by more than 1ºC (or
0.5ºC alternatively) in the servo
control mode. Flashes when skin
temperature 1 exceeds 40ºC in the
manual control mode.
Skin temperature probe alarm
indicator
Flashes when a wire of the skin
temperature probe has snapped or
short-circuited in the servo control
mode. Comes on when it has snapped
or short-circuited in the manual
control mode.
Baby check alarm indicator
Comes on in 15 minutes when the
heater output is set to 35% or higher.
Can be reset by pressing the alarm
silence/reset switch but comes on
again in another 15 minutes.
Canopy tilt alarm indicator
Comes on when the canopy is tilted.
Function
Symbol