Technical Data[12]
Main body
Power requirements Rating: AC120V; power consumption 700VA; frequency 60Hz
Operating voltage range: AC120V±10%
Classification Type of protection: Class 1 equipment
Degree of protection: Type BF applied part
Not for use in an air and flammable anesthetic gas mixture or an oxygen/
nitrous oxide and flammable anesthetic gas mixture
Mode of operation: Continuous operation
The Hi-Low stand is designed for intermittent operation. (Continuous opera-
tion of the pedal should be limited to one minute. Wait at least 20 minutes
before operating the pedal again.)
Operating conditions Ambient temperature: 18~30°C
Relative humidity: 30~75%
Atmospheric pressure: 70~106kPa
Storage conditions Ambient temperature: 0~50°C (32~122°F)
Relative humidity: 30~75%
Atmospheric pressure: 70~106kPa
Standards
IEC EN JIS
IEC60601-1 EN60601-1 JIS T0601-1
IEC60601-1-2 EN60601-1-2
IEC60601-2-21 EN60601-2-21 JIS T0601-2-21
Temperature control mode Manual control/servo control selectable
Heater capacity 500W
Skin temperature display range 30.0~42.0°C Accuracy: ±0.3°C
(86.0~107.6°F Accuracy: approx. ±0.5°F)
Skin temperature setting range 34.0~38.0°C (in 0.1°C increments)
(93.2~100.4°F (in 0.1°F increments))
Heater output setting range 0~100% (in 5% increments)
Heater output indication In 10 levels
Warming time indication Audible and visual alarms are given every 15 minutes of warming when
(Baby check alarm) the heater output is set to 35 % or higher in the manual control mode.
Timer 0~60 minutes 00 seconds (in one second increments)
Apgar mode/CPR mode selectable
Alarms Set temperature, baby check, power failure, system failure, skin temperature
probe, canopy tilt
Lighting LED lighting, intensity adjustable, 0~1200lux or more (center of the mat-
tress)
Mattress platform tilting Longitudinal tilt of ±13°(freely adjustable)
Communication connector RS-232C
Dimensions Main body: 84(W) x 110(D)cm Height 179~219cm (mattress surface 81~121cm
in height)
Mattress: 50(W) x 70(D) x 2.5(T) cm
Weight Approx. 105kg

Table of Contents