Accessories Skin temperature probe......................1
Mattress sheet .....................................1
Dust cover ............................................1
Operation Manual ...............................1
Pulse oximeter * ■
1,5,6,7
(Masimo)
SpO
2
•Measurement
Display range: 1~100%
Accuracy (in 70~100%): ±3 digits (during no motion conditions)*
2
±3 digits (during motion conditions)*
3
±3 digits (during low perfusion conditions)*
4
Update cycle: 1 second
•Alarmfunction
Setting range: Upper limit: OFF, 50~99% (in 1% increments)*
8
Lower limit: OFF, 45~95% (in 1% increments)*
8
Pulse
•Measurement
Display range: 25~240bpm
Accuracy (in 25~240bpm): ±3 digits (during no motion conditions)*
2
±5 digits (during motion conditions)*
3
±3 digits (during low perfusion conditions)*
4
Update cycle: 1 second
•Alarmfunction
Setting range: Upper limit: OFF, 80~240bpm (in 5bpm increments)*
8
Lower limit: OFF, 35~180bpm (in 5bpm increments)*
8
Sensor
Wavelength: 660nm (rd) 905nm (ir)
Brightness: ≤15mW
Non-sterile, latex-free
Mode
Averaging time: 4, 6, 8, 10, 12, 14, 16 seconds (selectable)*
8
(factory default: 8 seconds)
Sensitivity setting: APOD, NORM (normal), MAX (high)*
8
(factory default: MAX)
*1 This device is calibrated to display functional oxygen saturation.
*2 SpO
2 70~100% hypoxia was induced in healthy adult volunteers with light skin and dark skin. Accuracy
was checked in no motion conditions against a laboratory CO-oximeter and an ECG monitor. Variability
distribution at this accuracy was ±1SD. ±1SD covers 68% of all data.
*3 SpO2 70~100% hypoxia was induced in healthy adult volunteers with light skin and dark skin. Accuracy
was checked by 1~2cm tapping 2~4 times per second and by 2~3cm irregular tapping 1~5 times per second
against a laboratory CO-oximeter and an ECG monitor. Variability distribution at this accuracy was ±1SD.
±1SD covers 68% of all data.
*4 In a laboratory test, in the range of 70~100% SpO2, accuracy in low perfusion conditions was checked
against the Biotec Index2 Simulator and the Masimo Simulator, both of whose signal strength is 0.02% or
more and permeability 5% or more. Variability distribution at this accuracy was ±1SD. ±1SD covers 68%
of all data.
*5 The materials accessible to the patient and the user comply with ISO 10993-1.