Atom Infa Warmer i - Pulse Oximeter (Masimo) Specifications

Atom Infa Warmer i
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Accessories Skin temperature probe......................1
Mattress sheet .....................................1
Dust cover ............................................1
Operation Manual ...............................1
Pulse oximeter *
1,5,6,7
(Masimo)
SpO
2
•Measurement
Display range: 1~100%
Accuracy (in 70~100%): ±3 digits (during no motion conditions)*
2
±3 digits (during motion conditions)*
3
±3 digits (during low perfusion conditions)*
4
Update cycle: 1 second
•Alarmfunction
Setting range: Upper limit: OFF, 50~99% (in 1% increments)*
8
Lower limit: OFF, 45~95% (in 1% increments)*
8
Pulse
•Measurement
Display range: 25~240bpm
Accuracy (in 25~240bpm): ±3 digits (during no motion conditions)*
2
±5 digits (during motion conditions)*
3
±3 digits (during low perfusion conditions)*
4
Update cycle: 1 second
•Alarmfunction
Setting range: Upper limit: OFF, 80~240bpm (in 5bpm increments)*
8
Lower limit: OFF, 35~180bpm (in 5bpm increments)*
8
Sensor
Wavelength: 660nm (rd) 905nm (ir)
Brightness: 15mW
Non-sterile, latex-free
Mode
Averaging time: 4, 6, 8, 10, 12, 14, 16 seconds (selectable)*
8
(factory default: 8 seconds)
Sensitivity setting: APOD, NORM (normal), MAX (high)*
8
(factory default: MAX)
*1 This device is calibrated to display functional oxygen saturation.
*2 SpO
2 70~100% hypoxia was induced in healthy adult volunteers with light skin and dark skin. Accuracy
was checked in no motion conditions against a laboratory CO-oximeter and an ECG monitor. Variability
distribution at this accuracy was ±1SD. ±1SD covers 68% of all data.
*3 SpO2 70~100% hypoxia was induced in healthy adult volunteers with light skin and dark skin. Accuracy
was checked by 1~2cm tapping 2~4 times per second and by 2~3cm irregular tapping 1~5 times per second
against a laboratory CO-oximeter and an ECG monitor. Variability distribution at this accuracy was ±1SD.
±1SD covers 68% of all data.
*4 In a laboratory test, in the range of 70~100% SpO2, accuracy in low perfusion conditions was checked
against the Biotec Index2 Simulator and the Masimo Simulator, both of whose signal strength is 0.02% or
more and permeability 5% or more. Variability distribution at this accuracy was ±1SD. ±1SD covers 68%
of all data.
*5 The materials accessible to the patient and the user comply with ISO 10993-1.