EasyManua.ls Logo

Atom Infa Warmer i - Pulse Oximeter (Nellcor) Specifications

Atom Infa Warmer i
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
*6 All the sensors and patient cables that can be used with this device have been tested and verified by using
the Masimo/MX-1 module pulse oximeter technology.
*7 Since the measurements taken by the pulse oximetry equipment are statistically distributed, only about
two-thirds of them fall in the accuracy (Arms) range of values obtained with a CO-oximeter.
*8 Even if the power supply is interrupted due to power failure, disconnection of the power plug, or some
other cause, the preselected settings will be retained in the memory. When the power supply is resumed,
the unit will start operation with the last selected settings.
*9 The above specifications apply when the sensor and the patient cable specified by ATOM are used. For
information on the sensor and the patient cable specified by ATOM, contact your local Atom representa-
tive.
Pulse oximeter *
1,4,5
(Nellcor)
SpO
2
•Measurement
Display range: 1~100%
Accuracy (in 70~100%): ±3 digits (during no motion conditions)*
2
±3 digits (during low perfusion conditions)*
3
Update cycle: 2 seconds
•Alarmfunction
Setting range: Upper limit: OFF, 50~99% (in 1% increments)*
6
Lower limit: OFF, 45~95% (in 1% increments)*
6
Pulse
•Measurement
Display range: 25~240bpm
Accuracy (in 25~240bpm): ±3 digits (during no motion conditions)*
2
±3 digits (during low perfusion conditions)*
3
Update cycle: 2 seconds
•Alarmfunction
Setting range: Upper limit: OFF, 80~240bpm (in 5bpm increments)*
6
Lower limit: OFF, 35~180bpm (in 5bpm increments)*
6
Sensor
Wavelength: 660nm (rd) 900nm (ir)
Brightness: <15mW
EOG sterilized, latex-free
Mode
SatSeconds: OFF, 10, 25, 50, 100 (selectable)*
6
(factory default: OFF)
Fast response: OFF, ON*
6
(factory default: OFF)