Check the following points after connecting the power cord.(2)
Item to check Description
Mains on indicator The mains on indicator should come on.
Height adjustment mechanism
The mechanism should operate smoothly without making a strange
sound.
Operation panel
All the indicators on the operation panel, except for the power failure
alarm indicator and the battery operation indicator, should come on and
an audible alarm should sound when the power switch is turned on.
Skin temperature probe
An appropriate temperature should be displayed when the skin tempera-
ture probe is connected and the tip of the probe is held with your hand.
SpO
2 probe
SpO2 and pulse rate values should be displayed when the SpO2 sensor
is attached.
Various switches
Each switch should function properly when operated according to the
Operation Manual.
Lighting lamp
Intensity of the lighting lamp should increase gradually as the light control
knob is turned clockwise. The lighting lamp should go out when the light
control knob is turned fully counterclockwise.