Cate-
gory
Alarm name Condition causing the
alarm
Indication of the
alarm
Resetting the
alarm
Priority
(*1)
Alarm
silence
Pulse Oximeter (Masimo)
Incompatible
sensor
This alarm will occur if a non-
specified sensor or a non-spec-
ified patient cable is used.
“E03” will be dis-
played on the SpO
2
display. “–” will be
displayed on the
pulse rate display.
Replace the wrong
sensor or the wrong
patient cable with a
new one specified.
Warning
informa-
tion to
the user
-
Internal error
This alarm will occur in the
case of a self-diagnosis error,
a communication error, etc. of
the Masimo board.
“Err” will flash on
the SpO
2 display
and on the pulse
rate display.
Turn the power off.
If the alarm should
recur when the pow-
er is turned on again,
the unit is considered
to be defective.
Low
Switched
to warning
informa-
tion to the
user
Pulse Oximeter (Nellcor)
Pulse timeout
This alarm will occur if it be-
comes impossible to detect a
pulse during SpO
2 measure-
ment.
The pulse search
alarm indicator will
flash.
The alarm will be
reset automatically
when pulses are de-
tected normally.
High 2 min.
Pulse search
This alarm will occur when no
pulse is detected.
The pulse search
alarm indicator will
come on.
The alarm will be
reset automatically
when pulses are de-
tected normally.
Warning
informa-
tion to
the user
-
Sensor discon-
nect
This alarm will occur if the
patient cable or the sensor has
come off the unit.
Th e s en so r of f
alarm indicator will
flash
The alarm will be
reset automatically
when the sensor or
the patient cable is
connected properly.
Medium
Reset
(Turn the
function
off.)
Sensor off
This alarm will occur if the sen-
sor attached to the patient has
come off.
Th e s en so r of f
alarm indicator will
flash.
The alarm will be
reset automatically
when the sensor is
attached to the pa-
tient properly.
Medium 2 min.
Interference
This alarm will occur when
strong interference is detect-
ed.
The inter ference
alarm indicator will
come on or flash.
The alarm will be
reset automatically
when the noise has
decreased.
Warning
informa-
tion to
the user
-
Defective sensor
This alarm will occur if the sen-
sor is defective.
“E01” will be dis-
played on the SpO
2
display. “–” will be
displayed on the
pulse rate display.
Replace the defec-
tive sensor with a
good one. (When
the connector is re-
moved, Defective
sensor alarm will
be replaced with
“Sensor disconnect”
alarm.)
Medium 2 min.
Internal error
This alarm will occur in the
case of a malfunction, a soft-
ware error, a communication
error, etc.
“Err” will flash on
the SpO
2 display
and on the pulse
rate display.
Turn the pulse func-
tion off. (Actually, it
is impossible to reset
the alarm.)
Low
Switched
to warning
informa-
tion to the
user