26 INSTRUKCJA OBSŁUGI
____________________________________________________________________________________________________________________
Steps during calibration with external weight:
Press Menu key, choose Calibration
option with the cursor and press
ENTER.
Check if External load value matches
the value of external weight used for
calibration. If not, choose External
load option and enter correct value.
Choose External calibration option
with the cursor and press ENTER.
Wait until taring process is finished.
Put calibration weight matching
displayed value on.
Take off the calibration weight.
Wait until internal calibration is
finished.
The balance is ready to work.
USER FUNCTIONS&OPT.
Autotaring
PCS counting
. . .
Calibration
. . .
USER FUNCTIONS&OPT. \ CALIBRATION
1. External calibration
2. External load : 200g
3. Time of calibration : 2h
4. Temp. of calibration : 2.0
o
C
5. Report printout
6. Exit
Max ... Min ... e= ... d= ...
0.000 g
0% 100%
CALIBRATION
External calibration:
taring ...
CALIBRATION
External calibration:
Put external load _200.000 g
CALIBRATION
External calibration:
take off ext. load
CALIBRATION
Internal calibration:
please wait ...