Measurement Range Resolution Accuracy Response Time
O
2
0 to 20.9% 0.1% ±0.3% O
2
(on ue gas) T90 < 20 seconds
CO-H
2
0 to 10,000 ppm 1 ppm
±10 ppm (0 to 200)
±5% reading (201 to 2,000)
±10% reading (2,001 to 10,000)
T90 < 40 seconds
CO
HIGH
0 to 40,000 ppm 10 ppm
±10 ppm (0 to 100)
±10% reading (101 to 40,000)
T90 < 75 seconds
NO 0 to 3,000 ppm 1 ppm
±3 ppm (0 to 50)
±5% (51 to 2000)
±10% reading (> 2000)
T90 < 30 seconds
NO
2
0 to 500 ppm 1 ppm
±3 ppm (0 to 50)
±5 % (51 to 500)
T90 < 40 seconds
SO
2
0 to 5,000 ppm 1 ppm
±10 ppm ( 0 to 200)
±5 % reading (201 to 2,000)
±10 % reading (2,001 to 5,000)
T90 < 30 seconds
Approval Description
EN 50270:2015
EMC tested in accordance with European Directive 2014/30/EU.
Electromagnetic compatibility: electrical apparatus for the detection and
measurement of combustible gases, toxic gases and oxygen.
FCC: FCC Part 15.225 and RSS 210 – Intentional Radiator
IMPORTANT: When the PCA
®
400 is in Dilution Mode, the analyzer’s accuracy will be the
range stated in the following table plus an additional:
±1.0% O
2
,
±10% of reading for CO, CO, NO, NO
2
and SO
2
.
6.1.2 Approvals
6.1.3 Measurement Range & Accuracy