EasyManuals Logo
Home>Bante Instruments>Measuring Instruments>221

Bante Instruments 221 User Manual

Bante Instruments 221
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
35 I Dissolved Oxygen
Dissolved Oxygen
Calibration and Measurement
This section is applicable to model
821
900P/903P/904P
Prior to Use
Filling the Electrolyte Solution
1.1 Take out the dissolved oxygen electrode and electrolyte solution
from the packaging. Unscrew the membrane cap from the bottom
of the electrode, rinse the inside and outside with distilled water
and blot dry.
1.2 Fill the membrane cap halfway with electrolyte solution.
1.3 Screw membrane cap back onto the electrode. Some electrolyte
solution will overflow during this process.
1.4 Check the electrode, ensure that no air bubbles are trapped in the
electrolyte solution and the membrane is not creased or damaged.
Polarizing the Electrode
2.1 Insert the 6-pin connector into the connector socket on meter,
ensure the connector is fully seated.
2.2 Switch on the meter and wait 10 minutes for the electrode to
polarize.
The dissolved oxygen electrode must be polarized before use.
Remove air bubbles

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bante Instruments 221 and is the answer not in the manual?

Bante Instruments 221 Specifications

General IconGeneral
BrandBante Instruments
Model221
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals