10 µ S/cm, 84 µ S/cm, 1413 µ S/cm,
0 to 100° C (32 to 212° F)
Linear (0.0 to 10.0%/° C)
0.00 to 10.00 ppt (max. 200 ppt)
0.00 to 80.00 ppt, 0.00 to 42.00 psu
0 to 50° C (32 to 122° F)
60.0 to 112.5 kPa, 450 to 850 mmHg
0 to 105° C (32 to 221° F)
0 to 50° C (32 to 122° F)
0 to 60° C (32 to 140° F)
LCD, 80 × 60 mm (3.15 × 2.36 in.)
3 × 1.5V AA alkaline batteries or
5V DC power adapter
30 minutes after last key pressed
170 (L) × 85 (W) × 30 (H) mm
(6.69 × 3.35 × 1.18 in.)
Cause and Corrective Action
Electrode dried out.
Soak the pH electrode in 3M KCl solution for about
30 minutes.
Soak the ion selective electrode in 100 ppm standard
solution for about 30 minutes.
Soak the conductivity electrode in tap water for
about 10 minutes.
Measurement exceeded the maximum range.
Check the electrode and sample.
Drifting erratic
readings
Check whether electrode is contaminated, clogged
or broken.
pH buffer problem. Use freshly prepared buffer
solutions to calibrate the meter.
Electrode has expired. Replace pH electrode.
DO electrolyte solution is depleted.
Refilling electrolyte solution.
Zero oxygen solution is contaminated.
Replace the calibration solution.