EasyManua.ls Logo

Bausch OccluSense - Page 75

Bausch OccluSense
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
73
OccluSense
®
-Test-Sensor
Error (symptom) Cause Action
Handheld does not recognize the
OccluSense
®
-Test-Sensor
OccluSense
®
-Test-Sensor is not
properly inserted
Check position of the test sensor
in the handheld (chapter 12)
OccluSense
®
-Sensor
Error (symptom) Cause Action
The handheld does not recognize
the OccluSense
®
-Sensor.
OccluSense
®
-Sensor is not
properly inserted
The used OccluSense
®
-Sensor is
defective or manipulated
Check position of the test sensor
in the handheld and, if neces-
sary, reposition the sensor
(chapter 14.1.2)
Replace unrecognized Occlu-
Sense
®
-Sensor
Consult Bausch
(support@occlusense.com)
Treatment
Error (symptom) Cause Action
The handheld turns off after a few
recordings
There are no suitable recharge-
able batteries inserted in the
battery compartment
The batteries are defective
see manual (chapter 8.1)
Replace batteries
use of 3 x AAA NiMH recharge-
able batteries, 1000 mAh
(chapter 8)
The recordings are assigned to
another patient
A patient other than the patient
being treated is set active in the
OccluSense
®
-iPad-App
Create patient file in the Occlu-
Sense
®
-iPad-App (chapter 13.2)
Activate patient in the Occlu-
Sense
®
-iPad-App (chapter 14.2)

Table of Contents