73
OccluSense
®
-Test-Sensor
Error (symptom) Cause Action
Handheld does not recognize the
OccluSense
®
-Test-Sensor
• OccluSense
®
-Test-Sensor is not
properly inserted
• Check position of the test sensor
in the handheld (chapter 12)
OccluSense
®
-Sensor
Error (symptom) Cause Action
The handheld does not recognize
the OccluSense
®
-Sensor.
• OccluSense
®
-Sensor is not
properly inserted
• The used OccluSense
®
-Sensor is
defective or manipulated
• Check position of the test sensor
in the handheld and, if neces-
sary, reposition the sensor
(chapter 14.1.2)
• Replace unrecognized Occlu-
Sense
®
-Sensor
• Consult Bausch
(support@occlusense.com)
Treatment
Error (symptom) Cause Action
The handheld turns off after a few
recordings
• There are no suitable recharge-
able batteries inserted in the
battery compartment
• The batteries are defective
• see manual (chapter 8.1)
• Replace batteries
use of 3 x AAA NiMH recharge-
able batteries, 1000 mAh
(chapter 8)
The recordings are assigned to
another patient
• A patient other than the patient
being treated is set active in the
OccluSense
®
-iPad-App
• Create patient file in the Occlu-
Sense
®
-iPad-App (chapter 13.2)
• Activate patient in the Occlu-
Sense
®
-iPad-App (chapter 14.2)