72
Error (symptom) Cause Action
Battery compartment cannot be
closed
Battery compartment cover is
defective or missing
• Screwdriver is damaged or
missing
Consult Bausch
(support@occlusense.com)
The connection between Occlu-
Sense
®
iPad-App and handheld is
not possible.
Connect handheld with iPad App
(chapters 10)
Reset the handheld to factory
default (chapter 19.2)
No connection can be established
between iPad and handheld.
The WLAN configuration has been
set up incorrectly
Reset the handheld to factory
default (chapter 19.2)
• Connect iPad App to the handheld
(chapter 10) and re-enter
WLAN credentials
(chapter 10)
The recording of the bite force
distribution cannot be started
The registration process has not
been performed
The regularly requested function
test (with the test sensor) has
not been performed or it has
failed
Perform registration process,
see chapter 10.2.4
Perform or repeat function test
with test sensor (chapter 12)
Consult Bausch
(support@occlusense.com)
OccluSense
®
-Charging station
Error (symptom) Cause Action
The charging station cannot be
connected to the (local) power
supply.
The matching power supply plug
is defective or missing
Consult Bausch
(support@occlusense.com)
The handheld batteries are not
(correctly) charged.
There are no batteries in the
battery compartment
• The batteries were inserted
incorrectly
• The handheld is not correctly
positioned in the charging station
The charging station does not
recognize the inserted handheld
Insert batteries into the battery
compartment (chapter 8.1)
• Insert handheld into the charging
station (chapter 8.1)
Make sure that the handheld’s
position in the charging station
is correct and without any inter-
mediate layer