BENNING IT 115 Device description
- 22 -
4.2 Function selector switch
The function selector switch is intended for selecting the
test and measuring functions
"SETTINGS" mode
Key functions after having selected the test / measuring function:
UP / DOWN
Selects the sub-function of the adjusted test / measuring function
(only for rotary switch positions R
, Z
, Z
, FI/RCD)
Selects the parameters and limiting values
Start of measurement
Back / cancel
Key functions in the Parameters field:
Modifies the selected parameter
Selects the next parameter
Parameters and limiting values for evaluating the measuring results:
Parameter,
limiting value
No parameters / limiting values, indication: _ _ _.
ON
Measuring results – will be marked as "PASS" / "FAIL" according to
the parameters and limiting values set
Please find further information on how to use the test / measuring functions of the installation
tester in chapter 5. Measurements.