EasyManua.ls Logo

Benning IT 115 - 5.4.4 Automatic RCD Test

Benning IT 115
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BENNING IT 115 Device description
- 39 -
Figure 5.20: Example of a tripping current measurement
Results displayed:
I ........... tripping current
Uci ....... contact voltage at tripping current I or final value, if RCD does not trip
t ............ tripping time
5.4.4 Automatic test
The automatic RCD test is intended to perform a complete RCD test (tripping time at different
fault currents, tripping current and contact voltage) in a sequence of automatic tests controlled
by the installation tester.
Additional key
HELP / DISPLAY
As soon as measurement is finished, the "
HELP
" key toggles between
the upper and lower part of the result field.
How to perform an automatic test
Steps of the automatic test
Note
Select the FI/RCD function by means of the function selector
switch.
Set the sub-function to AUTO.
Set the testing parameters.
Connect the test cables to the test object (see figure 5.17).
Press the "
TEST
" key to start the measurement.
Start of test
Testing with I
N, 0 (step 1)
RCD should trip
Activating the RCD
Testing with I
N, 180 (step 2)
RCD should trip
Activating the RCD
Testing with 5I
N, 0 (step 3)
RCD should trip
Activating the RCD
Testing with 5I
N, 180 (step 4)
RCD should trip
Activating the RCD
Testing with ½I
N, 0 (step 5)
RCD must not trip
Testing with ½I
N, 180 (step 6)
RCD must not trip
Tripping current test, 0 (step 7)
RCD should trip
Activating the RCD
Tripping current test, 180 (step 8)
RCD should trip
Activating the RCD
End of test

Table of Contents

Other manuals for Benning IT 115

Related product manuals