8
ADVANCED USE MANUAL DD700
When a bit has a value of "1" this means that the corresponding signal is in the true condition; the
meaning of the signals is as follows:
<s1>
bit 0
minimum weighing operation signal
bit 1
blocked tare signal
bit 2
preset(1)/self-weighing(0) tare signal
bit 3
"center zero" signal
<s2>
bit 0
LSB weighing extension signal (only
ME)
bit 1
stable weight signal
bit 2
overload signal
bit 3
MSB weighing extension signal (only
ME)
<s3>
bit 0
entered tare signal
bit 1
deleted blocked tare signal (only
ME)
bit 2
invalid weight
bit 3
printing in progress
<s4>
bit 0
approved instrument
bit 1
failed converter
bit 2
error on scale configuration parameters
bit 3
not used
Available protocols: Cyclical, On request, ACK-NAK, With remote commands.