KBF / KBF-UL + KMF (E6) 09/2021 page 137/160
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them
and thus obtain a homogenous distribution of temperature and humidity.
23.4 KBF / KBF-UL Technical Data
Height, gross (incl. feet/castors)
Depth, gross (including door handle, con-
troller
triangle, connection and 30 mm for
cable)
mm / inch
730 / 28.74 880 / 34.65 970 / 38.19 1230 / 48.43
Wall clearance rear (minimum)
(spacer)
mm / inch
100 / 3.94 100 / 3.94 100 / 3.94 100 / 3.94
Wall clearance side (minimum)
Quantity of inner glass doors
Quantity of racks (regular)
Maximum permitted total load
Temperature data (without humidity)
Temperature range °C / °F
Temperature
uniformity (variation)
Climatic data (with humidity)
Temperature range °C / °F
Temperature
fluctuation
at 25 °C / 77 °F / 60% r.h.
at 40 °C / 104 °F / 75% r.h.
uniformity
at 25 °C / 77 °F / 60% r.h.
± K 0.2 0.3 0.2 0,2