9.7.4 Setpoint entry
Select the field Temperatureand enter the desired temperature setpoint.
KBF / KBF-UL setting range: -5 °C up to 70 °C, KMF setting range: -15 °C up to 100 °C.
Confirm entry with Confirm icon. The controller returns to the section view.
Select the field Humidityand enter the desired humidity setpoint.
KBF / KBF-UL setting range: 0% r.h. up to 80% r.h., KMF setting range: 0% r.h. up to 100% r.h.
Confirm entry with Confirm icon. The controller returns to the section view.
Select the field Fanand enter the desired fan speed setpoint.
Setting range: 40% up to 100% fan speed.
Confirm entry with Confirm icon. The controller returns to the section view.
9.7.5 Tolerance range
You can specify a temperature and humidity program tolerance range for each program section with differ-
ent values for the tolerance minimum and maximum. When the actual value exceeds the given threshold,
the program is interrupted. This is indicated on the display (see below). When the actual temperature is
situated again within the entered tolerance limits, the program automatically continues. Therefore, the du-
ration of the program may be extended due to the programming of tolerances.
Programming of tolerances may extend program duration.
An entry of “-99999” for the tolerance minimum means “minus infinite” and an entry of “999999” for the
tolerance maximum means “plus infinite”. Entry of these values will never lead to program interruption. The
entry of “0” for the tolerance minimum and/or maximum deactivates the respective tolerance function.
When requesting rapid value transitions, we recommend not programming tolerance values in order to
enable the maximum heating-up, cooling-down, humidification or dehumidification speed.