Binder VD 53 - Programming with Setting Step

Default Icon
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VD (E2.1) 07/2017 page 36/96
Program table corresponding to the diagram (with default setting “Ramp”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
Operation lines
O.LINE
S01
40
00:30
000
S02
40
00:01
000
S03
60
01:30
000
S04
60
00:01
000
S05
80
01:00
000
S06
80
00:01
000
S07
20
03:20
000
S08
20
00:01
000
You can now enter the values of such a program table to the RD3 program controller (chap. 8.2).
Program the end point of the desired cycle with an additional section (in our examples S05 for set-point
ramp and S08 for set-point step) with a section time of at least one minute. Otherwise, the program will
stop one section too early because the program line is incomplete.
8.1.2 Programming with setting “step”
With setting “Step” selected, you don’t need to program the transition section in the Program Editor.
With setting “step” the controller will equilibrate only to constant temperatures; programming
ramps is no longer possible.
The set-points are maintained constant for the duration of a program section. At the start of each program
section, the chamber heats up with the maximum speed in order to attain the entered set-point
Program entry as set-point step (example):
W/°C
t/min.
S01 S03 S05 S07
0
20
40
60
80
100
30
120
180
380