VD (E2.1) 07/2017 page 58/96
Maximum admissible load of the socket:
Standard 230 V / 50 Hz - 16 Amp
cUL-version 120 V / 60 Hz - 13 Amp
With CUL-version, connect only a UL listed vacuum pump with ratings of 120 V AC, 60 Hz,
and less or equal 12 Amp, 0.5 HP.
Connection to a vacuum source
Connect the vacuum connection (13) (small flange DN 16) at the
rear of the chamber at the top to a vacuum pump or to a domestic
vacuum system via a vacuum hose or a fixed vacuum pipe.
When using a vacuum hose, we recommend using the BINDER
connection kit VD Art. no. 8012-0146 (chap. 16.1
has an appropriate hose outlet at the back.
Figure 18: VD 53 with option “Vacuum module”, mounted, and
installed vacuum hose
CAUTION
Technical defects due to excessive negative pressure.
Danger of implosion.
Damage to the chamber.
∅ The end vacuum must NOT fall below the permissible level of 10
-2
mbar / 0.0003 inHg.
Install the pump according to th
e permissible end vacuum or limit the end vacuum via a
vacuum controller.
WARNING
Release of extracted vapors.
Danger of injury.
Damage to oven and pump by corrosion.
Remove the extracted vapors from the module via a suitable hose into e.g., a fume
extractor facility.
Connect the hose to the vacuum pump outlet.