For the protection of environment, do NOT throw wasted battery away. Please
inquire at the biomedical engineering lab of the hospital and dispose according to
the appropriate procedure to the authorized place (according to national
regulations).
Take care of the polarity when replacing the battery.
It is strongly recommended using the battery provided by the Bionet.
Using non-approved batteries may damage the equipment.
Effect of Lithium-ion Battery Technology on the Battery
This is explained about the lithium-ion battery technology.
The battery will discharge itself even though it is not connected to the equipment. Because
it is caused by the induced current in a circuit integrated with the lithium-ion battery, which
is own properties, it is self-discharged. The self-discharge rate increases by twice every 10°C
(18°F) increase in temperature.
Battery retention is more lost in the higher temperatures. Depending on battery life, the
volume of the full charge rate may decrease gradually or out of work. As a result, the total
amount of charge stored and available is reduced.
Inspection Instructions
Ensure battery performance by fully charging and discharging equipment every six months.
Storage Instructions
The battery must be stored between 20°C and 25°C (68°F and 77°F) from outside the
monitor. If the battery is connected to an AC power source in the system, the battery
temperature increases by 15°C to 20°C (59°F to 68°F) at room temperature and it shortens
the life of the battery
When AC power is connected in the monitor equipment, the battery doesn't work and
doesn't supply power. Battery life may be less than 12 months. Bionet recommends it should
be kept separate from the equipment during the moving.