V 1.31
2021. 05 28 / 171 Doc. # : BN-OP-C7
Battery Type: Lithium-ion battery
The lithium-ion battery is rechargeable batteries with lithium-ion cells. Each battery
cell contains an electrical residual measurement circuit and a safety circuit.
WARNING
- The battery charging display will show correctly if the battery is operating normally.
- Old or defective batteries are significantly reduced capacity or operating time
NOTE
The battery charging display will take up to 15 seconds to reflect the actual capacity of
the internal battery if AC power is not supplied.
NOTE
- Bionet recommends replacing the lithium-ion battery after 24 months of use.
- Battery life depends on usage.
Continuing battery use will reduce battery life and increase the frequency of
replacement.
- Recharge the battery after it is discharged to prevent pre-discharge.
Display Battery Power Status
: Battery Fully Charged
: Battery Charge Half-Full
: Battery Charge Low
: Battery Almost Fully Drained
Replacing Battery
When replacing battery for this device, the same type of battery should be used.
Type: Lithium-ion battery 3BL335-BIO-4(10.8V, 3250mAh)
When to replace: Battery will automatically be charged when the device is
connected to AC power and cannot be charged when separated from the device.
Battery is designed to have a charging cycle of 300 times or more. If the device only
lasts 20 minutes or less on battery power, the battery needs to be replaced.
Additionally, when a battery pack is damaged or leaking chemicals, replace it
immediately. Do not use damaged battery packs with the device.