Table 3.30 – Alternate Trigger Slope Mode Function Menu 2
Select slope trigger condition.
Turn the “Universal” knob to set the
slope time. Time setup range is 20ns-
10s.
Select the trigger level that can be
adjusted by “LEVEL” knob. You can
adjust “LEVEL A”, “LEVEL B” or adjust
them at the same time.
Enter “Trigger setup menu” (See Table
3.17).
Return to the first page of “Alternative
trigger menu”.
Operating Instructions:
To use alternate triggering, follow these steps:
1. Input two irrelative signals to channel 1 and channel 2.
2. Press the AUTO button.
3. Press the TRIG MENU button to enter “trigger menu”.
4. Press the “Type” option button to select “Alternative”.
5. Press the “Channels” option button to select “CH1-CH2”