Troubleshooting Guide
100
Chapter 7 Troubleshooting Guide
7.1 Frequently Asked Questions
Front Panel Operation
1. How do I exit the menu system?
- Press any of the top row of keys (i.e. DCV, ACV, etc.) or press Shift + Left Arrow key.
2. How do I see the data stored inside the buffer after running store reading function?
- Press Shift + Left Arrow key to enter the menu and use the Left or Right Arrow keys
until display shows D: SYS MENU.
- Press the Down Arrow key once, then use Left or Right Arrow keys again to 3: SAVED
RDGS.
- Press the Down Arrow key once more and you will see the stored readings. Use Left
or Right Arrow key to browse through all stored readings.
3. I cannot read AC or DC current.
- You might not be connecting to the correct inputs. For low current measurements (< 1
A), connect SENSE LO and INPUT LO terminals. For high current measurements (1
to 12 A), connect INPUT LO and 10 A terminals.
- Auto range is only available for low current measurements ( 12 mA and 120 mA(DCI)
ranges only). If you are measuring current outside of these two ranges, use the Up
Arrow key to manually select a higher current range.
- If you have connected SENSE LO and INPUT LO terminals with a current above 2 A,
the front panel protection fuse may have been blown already. Check the fuse and
replace it if necessary. See “3.3.2 Front Panel Fuse Replacement” for details.
4. How do I calibrate the instrument?
- Please contact B&K Precision for details.
5. How do I turn off the beep sound after every key presses?
- Press Shift + Left Arrow to go into the menu. Use the Left or Right Arrow keys until
display shows D: SYS MENU.
- Press the Down Arrow key once, then use Left or Right Arrow keys again to 8: KEY
SOUND.
- Press the Down Arrow key once more and the display will show SOUND: OFF. Use
Left or Right Arrow key to change it to ON.
- Press Auto to save the setting, and the key sound will now be disabled.
Remote Operation
1. How do I extract stored buffer readings over GPIB interface?
- Currently the command used for recalling buffer readings is only available for USB and
RS-232 interfaces.
2. I have connected the USB cable from the instrument to the computer, but it is not
working.
- The USB interface is a virtual COM. You first need to install the drivers,
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 5492B and is the answer not in the manual?

BK Precision 5492B Specifications

General IconGeneral
BrandBK Precision
Model5492B
CategoryMultimeter
LanguageEnglish

Related product manuals