SCPI Command Reference
92
6.3.5 SYSTem Subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in Table 6-5
Table 6-5 SYSTem Command Summary
Command
Function Description
Default
:SYSTem
:PRESET
:AZERo
:STATe <b>
:STATe?
:BEEPer
[:STATe] <b>
[:STATe]?
:LOCal
Return to system defaults
Path to set up autozero
Enable or disable autozero
Query autozero
Path to control beeper
Enable or disable beeper
Query state of beeper
Take multimeter out of remote state and restore front panel
operation
ON
ON
:PRESET
Command Syntax:
:SYSTem:PRESET
Description:
This command returns the instrument to factory defaults states.
:AZERo
:STATe <b>
Command Syntax:
:SYSTem:AZERo:STATe <b>
Command parameter:
<b> = 1 or ON Enable autozero
0 or OFF Disable autozero
Query:
:STATe? Query state of autozero.
Return String:
1 or 0
Description:
This command is used to enable or disable autozero. When enabled, accuracy is optimized. When
disabled, speed is increased at the expense of accuracy.
Note: Before you can enable or disable autozero, the multimeter must first be in the idle
state.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 5492B and is the answer not in the manual?

BK Precision 5492B Specifications

General IconGeneral
BrandBK Precision
Model5492B
CategoryMultimeter
LanguageEnglish

Related product manuals