Specifications
103
DC CHARACTERISTICS
CONDITIONS: SLOW or MED with filter count of 10
DC Voltage
Resolution, Full Scale reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Rate
Range
Resolution
Typical
Input
Impedance
Accuracy
(1 year)
Full Scale
Reading
Slow
120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 V
1 μV
10 μV
100 μV
1 mV
10 mV
>10G
>10G
>10G
10M ±1%
10M ±1%
0.02+0.008
(1)
0.01+0.004
(1)
0.01+0.004
0.01+0.004
0.01+0.004
119.999
1.19999
11.9999
119.999
1010.00
(2)
Med
120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 V
1 μV
10 μV
100 μV
1 mV
10 mV
>10G
>10G
>10G
10M ±1%
10M ±1%
0.02+0.015
(1)
0.01+0.008
(1)
0.01+0.008
0.01+0.008
0.01+0.008
119.999
1.19999
11.9999
119.999
1010.00
(2)
Fast
120.00 mV
1.2000 V
12.000 V
120.00 V
1000.0 V
10 μV
100 μV
1 mV
10 mV
100 mV
>10G
>10G
>10G
10M ±1%
10M ±1%
0.02+0.040
(1)
0.02+0.020
(1)
0.02+0.020
0.02+0.020
0.02+0.020
119.99
1.1999
11.999
119.99
1010.0
(2)
(1)
under REL status
(2)
1% over-range (1010 V) is readable at 1000 V range
Maximum input voltage: 1000VDC or peak ac for all ranges.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 5492B and is the answer not in the manual?

BK Precision 5492B Specifications

General IconGeneral
BrandBK Precision
Model5492B
CategoryMultimeter
LanguageEnglish

Related product manuals