SCPI Command Reference
99
:AUTO <b>
Command Syntax:
:TRIGger:DELay:AUTO <b>
Command parameter:
<b> = 1 or ON Enable auto delay
0 or OFF Disable auto delay
Query:
:AUTO? Query delay state
Return String:
1 or 0
Description:
This command is used for enable or disable trigger auto delay. If disabled, the trigger will delay
based on the delay time configured with “:TRIGger:DELay” command.
6.3.8 Common Commands
Common commands can be used in instruments that are SCPI-compliant. The common commands
available for the 5492B are described below:
*RST
Command Syntax:
*RST
Description:
Resets the instrument.
*TRG
Command Syntax:
*TRG
Description:
Sends a bug trigger to the instrument. Use this if trigger source is set to BUS.
*IDN?
Query Syntax:
*IDN?
Query return:
<product>,<version>,<sn number>
Example: 5492B Digital Multimeter, Ver1.0.00.00.01,123A45678
Description: Query the identification of the instrument.