Specifications
102
Chapter 8 Specifications
8.1 Technical Specifications
Specifications Assumptions
One year calibration cycle.
Operating temperature between 18°C to 28°C
Accuracy is expressed as: ±(% of reading + % of range ) after a 30 minute warm up and valid
for 10 PLC (slow)
Temperature Coefficient: add ± [0.1% × (the applicable accuracy)/ °C] for 0°C to 18°C and 28°C
to 40°C;
Relative Humidity: Up to 80% for 0°C to 28°C (75% RH for 10 MΩ and above ranges of
resistance measurement). Up to 70 for 28°C to 40°C.
Displayed Readings and Reading rates
Full Scale displayed reading
Reading rates on front panel (Reading/sec, Approximate)
2-Wire Ω (Ranges below 120 kΩ )
2-Wire Ω (120 kΩ range and above)
4-Wire Ω (Ranges below 120 kΩ)
4-Wire Ω (120 kΩ range and above)