Specifications
102
Chapter 8 Specifications
8.1 Technical Specifications
Specifications Assumptions
One year calibration cycle.
Operating temperature between 18°C to 28°C
Accuracy is expressed as: ±(% of reading + % of range ) after a 30 minute warm up and valid
for 10 PLC (slow)
Temperature Coefficient: add ± [0.1% × (the applicable accuracy)/ °C] for 0°C to 18°C and 28°C
to 40°C;
Relative Humidity: Up to 80% for 0°C to 28°C (75% RH for 10 and above ranges of
resistance measurement). Up to 70 for 28°C to 40°C.
Displayed Readings and Reading rates
Full Scale displayed reading
Slow
Med
Fast
119,999
119,999
11,999
Reading rates on front panel (Reading/sec, Approximate)
Measurement Function
Slow
Med
Fast
DCV
4
16
57
DCA
4
16
57
ACV
3
4
25
ACA
3
4
25
2-Wire Ω (Ranges below 120 kΩ )
4
16
57
2-Wire Ω (120 kΩ range and above)
4
16
25
4-Wire Ω (Ranges below 120 kΩ)
3
10
33
4-Wire Ω (120 kΩ range and above)
3
10
20
Freq
1
————
————
Period
1
————
————
Diode
————
16
————
Continuity
————
————
57
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 5492B and is the answer not in the manual?

BK Precision 5492B Specifications

General IconGeneral
BrandBK Precision
Model5492B
CategoryMultimeter
LanguageEnglish

Related product manuals