Specifications
108
FREQUENCY AND PERIOD CHARACTERISTICS
CONDITIONS: SLOW RATE (GATE TIME 1 Sec)
Frequency
Accuracy: ± (% of Reading)
5 ~ 10 Hz
10 ~ 100Hz
100 ~ 100 kHz
100 k ~ 1 MHz
(1)
200 mVrms
40 mVrms
40 mVrms
100 mVrms
(1)
If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is
guaranteed.
(2)
Specified accuracy at input >5 % of range.
Period
10 μs ~ 10 ms
10 ms ~ 100 ms
100 ms ~ 200 ms
40 mVrms
40 mVrms
200 mVrms
(1)
If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is
guaranteed.
(2)
Specified accuracy at input >5 % of range.
Maximum Crest factor: 3.0 at full scale
Maximum input voltage: 750Vrms
≤3×10
7
Volt-Hz for all ranges
Input impedance: 1 MΩ ± 2% in parallel with capacitance <100pF
Maximum DCV: 500 V for all AC ranges
Triggering and MemoryTrigger Delay: 0 to 6000ms (1 ms step size)
Memory: 512 readings can be stored
Math Functions
Rel, Max/Min/Average/StdDev (for stored readings), dB, dBm, Limit Test, % and mX+b
dBm Reference Resistance: 1 Ω to 9999 Ω (1Ω step size)