EasyManua.ls Logo

BK Precision 5492B - CALCulate3 Subsystem (Limit Test); :LIMit[1]:UPPer Command (Set Upper Limit); :LIMit[1]:LOWer Command (Set Lower Limit)

BK Precision 5492B
111 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCPI Command Reference
77
:DATA?
Command syntax:
:CALCulate2:DATA?
Return String:
<reading><sp>
where
<reading> - This is the stored reading and consists of 8 characters, which includes decimal.
If the reading does not take up 8 characters, the unused characters become spaces.
For example: If the stored reading is 11.0016, the return string is: _ 11.0016
If the stored reading is 0.326, the return string is: _ _ _0.326
(where _ is a space)
<sp> - This is a space character.
Description:
This query command will return a statistical calculation result based on the stored readings inside
the buffer. It can be either the Maximum reading, minimum reading, the average of the stored
readings, or the standard deviation based on the stored readings. The “:CALC2:FORM” command
is used to select which calculation result to return after this query.
:CALCulate3 Subsystem
These commands are used to configure and control the limit test.
:LIMit[1]:UPPer <n> Specify upper limit
:LIMit[1]:LOWer <n> Specify lower limit
Command syntax:
:CALCulate3:LIMit[1]:UPPer <n>
:CALCulate3:LIMit[1]:LOWer <n>
Command Parameter:
<n> = -100e6 to 100e6 Specify limit value
Query:
:UPPer? Query upper limit
:LOWer? Query lower limit
Return String:
<n>
where
<n> - Specified limit value.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents