SCPI Command Reference
89
:STATe <b>
Command syntax:
[:SENSe[1]]:CURRent:AC:AVERage:STATe <b>
[:SENSe[1]]:CURRent:DC:AVERage:STATe <b>
[:SENSe[1]]:VOLTage:AC:AVERage:STATe <b>
[:SENSe[1]]:VOLTage:DC:AVERage:STATe <b>
[:SENSe[1]]:RESistance:AVERage:STATe <b>
[:SENSe[1]]:FRESistance:AVERage:STATe <b>
Command parameter:
<b> = 1 or ON Enable the digital filter
0 or OFF Disable the digital filter
Query:
:STATe? Query state of digital filter
Return String:
1 or 0
Description:
These commands are used to enable or disable the digital filter for the current measurement
function. When enabled, readings will be filtered according to how the filter is configured.
:TCONtrol <name>
Command syntax:
:CURRent:AC:AVERage:TCONtrol <name>
:CURRent:DC:AVERage:TCONtrol <name>
:VOLTage:AC:AVERage:TCONtrol <name>
:VOLTage:DC:AVERage:TCONtrol <name>
:RESistance:AVERage:TCONtrol <name>
:FRESistance:AVERage:TCONtrol <name>
Command parameter:
<name> = REPeat Select repeating filter
MOVing Select moving filter
Query:
:TCONtrol? Query filter type
Return String
REP or MOV
Description:
These commands are used to select the type of averaging filter for the specified function.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com