Weighing Terminal EWI / MWIPassword protected menu (advanced functions) | 6
6.2.3 Password protected menu P08 - overview CAL
If the calibration was not successful, the error message ."Error" is displayed.
Repeat the entire process as a failure may have prevented a successful
calibration. After calibration, the scale should be checked for correct calibration
and linearity. Repeat the process if necessary, and make sure that the scale is
stable before you load a weight.
If you want to discard the calibration, press:
Note
Acclimatisation of the weighing system to the environment (min.10 minutes) is
necessary for successful calibration. Make sure there are no objects on the
weighing plate.
CAL
SPAn
LInEAR
nU-U
GrAUTy
CALSL
CAL
Linearity
mV/V-value
Gravity
Calibration selection
CoDE
Menu navigation
SetIn9
1x
P08
P08-01
P08-02
P08-03
P08-04
P08-05
P08