Page 24
BOSCHE Weighing Systems
Version 4.0
2024-03-27
4 | Operation Weighing Terminal EWI / MWI
4.2 Error message
If no digitizing unit is recognized after switching on and the display self-test, the
following message appears in the display:
Please check if a scale is connected. For further error messages see chapter 7
“Troubleshooting“.
4.3 Power zero limits
Legal for trade: Initial zero setting 20 % of the weighing range.
Not legal for trade: Initial zero setting +/- 100 % of the weighing range.
If the scale is switched on outside the power zero limits the weight value actu-
ally contained on the scale is displayed after you have confirmed with the Tare
Button the error message.
If the load is on the scale exceeds the weighing capacity - OL - appears on the
display.
Note
At noncertified scales the initial zero can be set later in the menu.