How to Use the Clearway 2 46
DOC-001000 Ver 6 Clearway 2 User Manual .
The battery symbol will change depending on the state of the battery.
Internal battery Connected
Battery Charging
Battery Not Connected
Battery Fault. Contact your provider
Battery Under/Over Temperature
<10°C or >45°C
The battery LEDs on the front of the Clearway 2 will be solid if
the battery is in use, and flash if the battery is charging.
Internal Battery
If available, an internal battery can be fitted inside the Clearway 2 (see Back
Panel, section 3.4). The battery pack is not designed to be replaced by the user.
If you experience improper function or replacement is required, contact your
homecare provider, or Breas.
An internal battery will run for: >30 minutes in MI-E modes (based on settings
of +40/-45 cmH
2
O, and Ti, Te, and Pause times of 2 seconds). The capacity
of lithium-ion batteries degrades over time and use, therefore older batteries
and/or those that have been used extensively may not deliver the same running
times as stated. Users that require battery back-up should regularly monitor the
run-time they are able to achieve and replace the battery when necessary.
External Battery
If available, the external battery may be connected to the Clearway 2 using the
external battery connector at the rear (see Back Panel, section 3.4). An external
battery may be connected and disconnected as required.