278580 Revision A 123
LF80C ZeniD RFID Reader 7 Operation
Product Manual 7.3 Operation of SECS Protocol
Message S1F15 sets the reader offline
Host to Reader: S1F15
10:32:53 Outgoing: ENQ ( 05 )
10:32:53 Incoming: EOT ( 04 )
10:32:54 Outgoing: Length Byte ( 0A )
10:32:54 Outgoing: Header ( 01 FF 81 0F 80 01 00 00 00 02 )
10:32:54 Outgoing: Checksum ( 13 02 )
10:32:54 Incoming: ACK ( 06 )
Reader to Host: S1F16
10:32:54 Incoming: ENQ ( 05 )
10:32:54 Outgoing: EOT ( 04 )
10:32:54 Incoming: Length Byte ( 0D )
10:32:54 Incoming: Header ( 81 FF 01 10 80 01 00 00 00 02 )
10:32:54 Incoming: Data ( 21 01 00 )
10:32:54 Incoming: Checksum ( 02 36 )
10:32:54 Outgoing: ACK ( 06 )
Message S1F17 sets the reader online
Host to Reader: S1F17
10:33:24 Outgoing: ENQ ( 05 )
10:33:24 Incoming: EOT ( 04 )
10:33:24 Outgoing: Length Byte ( 0A )
10:33:24 Outgoing: Header ( 01 FF 81 11 80 01 00 00 00 04 )
10:33:24 Outgoing: Checksum ( 17 02 )
10:33:24 Incoming: ACK ( 06 )
Reader to Host: S1F18
10:33:24 Incoming: ENQ ( 05 )
10:33:24 Outgoing: EOT ( 04 )
10:33:24 Incoming: Length Byte ( 0D )
10:33:24 Incoming: Header ( 81 FF 01 12 80 01 00 00 00 04 )
10:33:24 Incoming: Data ( 21 01 00 )
10:33:24 Incoming: Checksum ( 02 3A )
10:33:24 Outgoing: ACK ( 06 )