EasyManua.ls Logo

Cardinal 212 Series - Page 74

Cardinal 212 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
212/212X Installation & Technical
8200-M639-O1 Rev D 09/14
Page 66
cells= (Number of Load Cells)
5. Press the ENTER key to show the current setting.
6. If the setting displayed is acceptable, press the ENTER key
again to save it and advance to the Sio setup prompt.
7. Otherwise, using the numeric keys enter a new setting and
then press the ENTER key.
Allowable settings for CELLS= are 1 to 10.
8. Starting at the left and proceeding right, a series of dashes will
appear on the display.
9. Next, starting at the left and proceeding right, the dashes will
disappear, after which the display will show: SIO?.
IMPORTANT! When calibrating using mode 1, the empty
scale weight is established in order to determine when the
scale is over capacity. If calibrating with weight on the
scale is necessary, we recommend that you later calibrate
again once the scale is empty in order to establish a
correct empty scale weight.

Table of Contents

Related product manuals