CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com 39
23.6.1.5 Command: U<CR> (55h 0dh)
Response: Changes units of measure (simulate UNIT key) and return scale status with new units,
The new measure unit should be allowed to use
<LF> U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>
23.6.1.6 Command:
L<CR> (4ch 0dh)
Response: If Hold function can be activated, it will enable/disable hold function (simulate HOLD key),
and returns scale status.
<LF> H
1
H
2
H
3
H
4
<CR><ETX>
23.6.1.7 Command: X<CR> (58h 0dh)
Response: power off the scale, just like press down the ON/OFF key to turn off the scale.
23.6.1.8 Command:
all others
Response: Unrecognized command
<LF>? <CR><ETX>
23.6.2 Summary of Command and Response:
Command
Response
ASCII HEX
W<CR> 57 0d
Read scale weight:
<LF>^^^^^^^^U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---over capacity
<LF>________U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---under capacity
<LF>-------- U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---zero-point error
<LF><p>W
1
W
2
W
3
W
4
W
5
<dp>W
6
U
1
U
2
U
3
U
4
U
5
<CR><LF>H
1
H
2
H
3
H
4
<CR><ETX>
---normal data
S<CR> 53 0d <LF> H
1
H
2
H
3
H
4
<CR><ETX>; read scale status
Z<CR> 5a 0d <LF> H
1
H
2
H
3
H
4
<CR><ETX> ; simulate ZERO key
T<CR> 54 0d <LF> H
1
H
2
H
3
H
4
<CR><ETX> ; simulate TARE key
U<CR> 55 0d <LF> U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>; simulate UNIT key
L<CR> 4c 0d <LF> H
1
H
2
H
3
H
4
<CR><ETX>; simulate HOLD key
X<CR> 58 0d power off the scale, simulate OFF key
others
<LF>? <CR><ETX>