5. SET-UP
5-1. Calibration
Adjust weight balance between Real weight” on the load cell(Weight Part) and “Displayed weight of
Indicator”. When you replace LOAD CELL or Indicator, you have to do Calibration process once again
5-2. Test Weight Calibration
Prepare At least 10% of Max. capacity of your weighing scale
Step 1. Enter Calibration Mode
Turn on the Power + 3 KEY display
Press 3 KEY key display
Press ENTER key to start “Calibration Mode”.
Remarks : Go to next step with save ENTER key / Back to previous step CLEAR key
Step 2. Digit(Division)setting ( display)
Press 0 KEY and stop the optimal Digit value.
(Whenever pressing 0 KEY, digit value will be changed, like 01-02-05-10-20-50)
Press ENTER KEY and save change and move to next step.
TEST
SET-CAL
d _1