2-2. Calibration Process
When Max.capacity/digit
value is over 20.00
Re-input the Max. Capacity, less than 20.00
(Max. Capacity / Digit)
Standard weight value is
over than Max. Capacity
Re-input Standard weight value with Number
keys, under Max. Capacity
Standard weight value is
less than 10% of Max.
Capacity
Re-input Standard weight value with Number
keys, more than 10% of Max. Capacity
1. Amp. Gain is too big
2. Sig+ and Sig- wire
connection error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too small)
1. Amp. Gain is too small
2. Sig+ and Sig- wire
connection error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and real
weight, please re-input the value
(set value is too big)
Under “F-function” model,
set value is “N.A”
Check the correct value and re-input
When there is continuous
vibration on the weighing
part,, indicator can not
process calibration any
more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting condition
check